Year : 2012  |  Volume : 3  |  Issue : 1  |  Page : 46-51  

An alternative subcoronary implantation technique decreases the risk of complete heart block after stentless aortic valve replacement


1 Department of Cardiothoracic Surgery, Changhai Hospital, The Second Military Medical University, Shanghai, China
2 Swedish Heart and Vascular Institute, Swedish Medical Center, Seattle, WA, USA
3 Division of Cardiac Surgery, The University of Alberta, Edmonton, Canada

Date of Web Publication9-Jan-2012

Correspondence Address:
Shaohua Wang
2H2.34 Walter Mackenzie Centre, 8440 - 112th Street, Edmonton, AB - T6G 2B7
Canada
Login to access the Email id


DOI: 10.4103/0975-3583.91594

PMID: 22346147

Get Permissions

   Abstract 

Context: Stentless aortic valve prostheses have excellent hemodynamic profiles, but may be associated with an increased postoperative heart block when compared to conventional prostheses. Aims: To assess the effect of an alternative subcoronary implantation technique on postoperative complete heart block (CHB) and permanent pacemaker implantation (PPI) following aortic valve replacement (AVR) with stentless tissue valves. Settings and Design: A total of 130 consecutive patients undergoing AVR with stentless tissue valves by a single surgeon were studied retrospectively. Materials and Methods: A stentless tissue valve was implanted into 80 patients using the conventional modified subcoronary implantation technique, and 50 patients received a stentless tissue valve by an alternative subcoronary technique in which the inflow suture line is raised at the level of right-non coronary commissure. Data were collected at the time of hospital discharge and at 6-12 months postoperatively. Statistical Analysis Used: Independent samples t-test was used to compare continuous variables, and categorical variables were compared with the chi-square test. Results: Use of this new method reduced postoperative CHB (4.0% vs. 16.3%, P = 0.033), with fewer patients requiring PPI in the early postoperative period (6.0% vs. 18.8%, P = 0.041). Echocardiographic examination showed no differences in the effective orifice area, peak and mean aortic valve gradients, or left ventricular mass index between groups. Trivial or mild aortic regurgitation was found in 3.9% of patients with the conventional modified technique and 4.0% of patients undergoing the alternative technique, during follow-up (P > 0.05). Conclusions: Compared to the conventional subcoronary implantation technique, the alternative subcoronary implantation technique reduces the incidence of new CHB and the requirement of PPI following AVR with stentless tissue valves while preserving mid-term hemodynamic function.

Keywords: Aortic valve replacement, heart block, pacemaker, stentless aortic valve


How to cite this article:
Song Z, Lehr EJ, Wang S. An alternative subcoronary implantation technique decreases the risk of complete heart block after stentless aortic valve replacement. J Cardiovasc Dis Res 2012;3:46-51

How to cite this URL:
Song Z, Lehr EJ, Wang S. An alternative subcoronary implantation technique decreases the risk of complete heart block after stentless aortic valve replacement. J Cardiovasc Dis Res [serial online] 2012 [cited 2013 Apr 19];3:46-51. Available from: http://www.jcdronline.com/text.asp?2012/3/1/46/91594


   Introduction Top


Cardiac conduction defects are well-known complications of aortic valve surgery. Mechanical trauma to the conduction system is the most common contributing factor. Higher degree atrioventricular block (AVB), although often reversible, may also necessitate permanent pacemaker implantation (PPI). It has been reported that the incidence of conduction defects requiring PPI in patients who underwent aortic valve replacement (AVR) with stented prosthetic valves ranges from 3 to 6%. [1],[2],[3] However, the need for PPI after AVR with stentless tissue valves is not well characterized.

Stentless tissue valve prostheses are most commonly implanted using the conventional modified subcoronary implantation technique. In this position, there is a risk of placing inflow sutures in the atrioventricular (AV) conduction system below the commissure between the right and non-coronary cusps [Figure 1]a. We previously reported that patients undergoing AVR with stentless tissue valves have increased need for PPI in the early postoperative period. [4] Therefore, we improved our technique for placing the inflow suture line when implanting the stentless tissue valve in the subcoronary position. We compared the incidence of new PPI insertion in patients receiving stentless tissue valves in the subcoronary position using two different implantation techniques. Additionally, the hemodynamic performance of the stentless tissue valve implanted using the new technique was assessed.
Figure 1: (a) The conventional modified subcoronary implantation technique. Inflow sutures are placed in a continuous fashion in a horizontal plane at the level of the nadir of the attachments of the native aortic valve leaflets to the aortic annulus. (b) The alternative subcoronary implantation technique. The inflow suture line is elevated at the right-non coronary commissure to minimize the risk of injuring the conduction system

Click here to view



   Materials and Methods Top


Patients

All adult patients who underwent AVR with a stentless tissue valve (Freestyle® Heart Valve, Medtronic, Minneapolis, MN, USA) by a single surgeon between January 2001 and June 2009 were considered for inclusion in this retrospective study. Exclusion criteria were the presence of preoperative PPI or automated implantable cardioverter-defibrillator, concomitant other valve replacement, aortic root replacement, aortic root enlargement procedure or left ventricular aneurysmectomy, patients receiving allograft or autograft (Ross procedure) valves, previous or concurrent septal myectomy and concomitant maze procedure. Based on these criteria, we retrospectively reviewed the charts for all patients who received stentless tissue valves over the course of the study period. Patients were divided into two groups according to the implantation technique. The conventional modified subcoronary implantation technique was used in 80 patients and 50 patients underwent AVR using the alternative subcoronary implantation technique. Concomitant procedures included mitral valve repair and coronary arterial bypass grafting. Data collection was completed at the time of hospital discharge and at follow-up (97.7% complete at 6-12 months postoperatively). Patient records from each group were analyzed for clinical characteristics, electrocardiograms (ECGs), echocardiographic data, indication for surgery, postoperative complications, the development of new complete heart block (CHB) and the need for PPI in the early postoperative period.

Electrocardiogram

Twelve-lead resting ECGs were recorded on a Hewlett-Packard electrocardiograph with a 0.05-100 Hz filter before and after operation. QRS duration, QT and QTc intervals were determined directly by the built-in-computer software. Experienced cardiologists performed analysis of the ECGs.

Echocardiographic assessment

Echocardiographic assessment included M-mode and Doppler studies preoperatively and at 6-12 months postoperatively. Left ventricular ejection fraction (LVEF), left ventricular wall thickness, peak and mean aortic gradients, effective orifice area, and left ventricular mass were determined preoperatively and on follow-up examination. Prosthetic valve competence was assessed on the follow-up echocardiogram.

Surgical techniques

All operations were performed under standard cardiopulmonary bypass. Myocardial protection was achieved by direct infusion of cold hyperkalemic blood cardioplegia into the coronary ostium. A transverse aortotomy provided exposure of aortic valve. After excision of the aortic valve and annular debridement, the aortic annulus was sized. Stentless valves were inserted in the subcoronary position using the conventional modified subcoronary technique as previously described [5] (n = 80) or using our alternative technique (n = 50). In the conventional modified subcoronary technique, the inflow suture line was constructed using three continuous 4-0 polypropylene sutures placed below each commissure in a horizontal plane based on the level of the nadir of the attachments of the native valve leaflets to the aortic valve annulus [Figure 1]a. In the alternative technique, the inflow suture line was constructed in the same horizontal plane, but was raised at the level of right coronary commissure to avoid the conduction system [Figure 1]b.

Criteria for permanent pacemaker implantation

Patients with symptomatic second-degree AVB or third-degree AVB persisting more than 7 days postoperatively were considered for PPI according to the ACC/AHA/NASPE 2002 guidelines for the implantation of cardiac pacemakers and anti-arrhythmia devices. [6] An attending electrophysiologist determined the requirement for PPI.

Statistical analysis

Statistical analysis was performed using SPSS 13 for windows (SPSS Inc., Chicago, IL, USA). Data were expressed as mean value ± standard deviation and assumed a normal distribution for continuous variables and as numbers with percentage for categorical variables. Independent samples t-test was used to compare continuous variables, and the chi-square test was used to compare categorical variables. Results were assessed with a 95% confidential interval (CI) and P values <0.05 were considered to be significant.

Our institutional Health Research Ethics Board approved the study protocol and waived the requirement for individual patient consent. The authors had full access to the data and took responsibility for their integrity. All authors have read and agreed to the manuscript as written.


   Results Top


The clinical characteristics of the study patients are summarized in [Table 1] which outlines the clinical demographics of the two groups. Patients in both groups were homogenous with respect to age, gender, concomitant disease, preoperative diagnosis, left ventricular function, ECG and echocardiography characteristics.
Table 1: Preoperative clinical and demographic characteristics for patients undergoing stentless aortic valve replacement

Click here to view


Relevant operative and postoperative data are detailed in [Table 2]. There were three perioperative deaths, two patients (2.5%) in the conventional group and one patient (2.0%) in the alternative group (P > 0.05). No differences in the incidence or type of concomitant operation and postoperative events were identified between groups. Aortic cross-clamp times were similar with both surgical techniques (P > 0.05).
Table 2: Operative and postoperative data of patients undergoing stentless aortic valve replacement

Click here to view


Use of the alternative subcoronary implantation technique dramatically reduced the development of CHB in the early postoperative period compared to patients who underwent conventional stentless aortic valve implantation (4.0% vs. 16.3%, P = 0.033). This reduction in CHB translated into a reduced need for PPI in the alternative implantation technique group (6.0% vs. 18.8%, P = 0.041).

Echocardiographic parameters were favorable in both stentless valve groups. Altering the subcoronary implantation technique had no impact on the effective orifice area, peak aortic gradient, mean aortic gradient, LVEF or left ventricular mass index. No patient in either group had clinically significant postoperative aortic regurgitation during follow-up. Trivial or mild aortic regurgitation occurred in 2 (4.0%) of patients who had stentless tissue valve replacement with the alternative subcoronary technique and in 3 (3.8%) patients who underwent the conventional technique. On follow-up echocardiography, there was no difference in the number of patients with aortic insufficiency or patient prosthesis mismatch between groups [Table 3].
Table 3: Postoperative echocardiographic data of patients during follow-up

Click here to view



   Discussion Top


AVR is recommended for patients with symptomatic severe aortic valve disease. Debate exists regarding the selection of the best valve prosthesis to enhance surgical and long-term outcomes. As compared to traditional stented tissue valves, newer stentless tissue valves have been associated with early left ventricular remodeling and improvement in left ventricular function while providing superior hemodynamics. [7],[8]

AV conduction disturbances requiring PPI have been a well-known complication of AVR since the introduction of the cardiac valve prosthesis in the early 1950s. Injury to the AV conduction pathways during AVR by annular decalcification, suture placement and pressure on the conduction tissue can cause heart block requiring postoperative PPI. [9],[10] AVB generally results from trauma to the AV node or His bundle in the region of the membranous septum and right trigone beneath the non-coronary cusp-right coronary cusp commissure. PPI is required in approximately 6% of patients after isolated AVR and in up to 6.5% of patients after AVR with concomitant coronary artery bypass. [6],[7],[8],[9] Patient-, surgeon- and valve-specific risk factors for AVB following AVR have been described. Bicuspid aortic valve, annular calcification, hypertension, preexisting conduction disturbances, previous myocardial infarction and total perfusion time are significant predictors of PPI postoperatively. [2],[3],[11],[12] Totaro et al. [13] demonstrated that a continuous inflow suture line rather than interrupted sutures is associated with postoperative AV conduction defects and PPI. [13] Interestingly, it was recently reported that the incidence of heart block was strikingly higher in patients randomized to aortic root replacement with a Freestyle® valve compared to an allograft valve (9% vs. 0%, P < 0.01). The authors did not suggest any reasons for this finding, [14] but it is possible that the more rigid sewing ring on the Freestyle® valve compared to a homograft valve exerts more pressure on the conduction tissue. Therefore, the Freestyle® valve may necessitate implantation further away from the conduction tissue, as we describe herein, than is necessary when implanting a homograft valve.

Limited data have been presented describing the development of irreversible AVB requiring PPI after AVR with stentless tissue valves. Recent studies reported the incidence of PPI following stentless tissue valve replacement to be as high as 8.3-19%. [15],[16],[17] We also previously reported that subcoronary stentless AVR is an independent predictor of postoperative heart block and the requirement for permanent cardiac pacing following AVR. [4]

During stentless valve implantation using the conventional modified subcoronary technique, the inflow sutures are placed below each commissure so that the suture line sits in a horizontal plane at the level of the nadir of the attachments of the native aortic valve leaflets to the aortic valve annulus. Consequently, the AV conduction system is liable to be injured when the inflow suture line is constructed beneath right-non commissure. In the current series, 16.3% of patients undergoing stentless tissue valve implantation by the conventional modified subcoronary technique developed CHB in the early postoperative period.

To minimize the risk of trauma to the conduction system, we improved the conventional modified subcoronary technique by raising the inflow suture line at the level of the commissure of the right and non-coronary leaflets. Using this alternative subcoronary implantation technique reduced the incidence of postoperative CHB and PPI back to an incidence in keeping with AVR with conventional valves. [12]

When altering the position of the inflow suture line, care must be taken to avoid distortion of the prosthetic valve, which could adversely impact the hemodynamics and competency of the valve. In this series, peak and mean gradients were low in the alternative subcoronary technique group and were not different from stentless valves implanted with the conventional technique. Postoperative echocardiography in patients following stentless tissue valve replacement using alternative subcoronary technique identified 4.0% of patients with trace/mild aortic regurgitation; these outcomes are in accordance with previous reports [18] for stentless valve implantation. Our results suggest that the alternative subcoronary implantation technique does not adversely impact the hemodynamic performance of stentless tissue valves and can provide excellent short-term clinical results. With the inflow suture line moved closer to the coronary ostia, care must also be taken to avoid compromising coronary flow. In this study, no coronary occlusion or myocardial ischemia occurred in any patient (data not shown).

As with all retrospective studies, our analysis may be subject to inherent bias. To reduce selection bias, all patients undergoing AVR by a single surgeon during the study period were considered for inclusion in this study. Although patients were enrolled consecutively and were comparable, they were not randomized. It is also possible that learning curve issues or a bias from using historical controls contributed to the differences in PPI between the two techniques. However, PPI was evenly distributed across patients in all time points throughout patients undergoing the conventional modified subcoronary technique and PPI rate fell immediately upon switching to the new technique. A much larger cohort investigation is required to confirm our findings regarding the alternative subcoronary implantation technique. Long-term studies are required to confirm excellent valve function at greater than 10 years.

Results from this study indicate significantly reduced requirement for PPI with the alternative method of subcoronary implantation of the stentless tissue valve. No adverse effect was determined by altering the position of the inflow suture line, and short to mid-term hemodynamic function of stentless tissue valves inserted using this technique remains excellent.


   Acknowledgment Top


We are indebted to Mary Ann James for her assistance in data collection and to Bin Liang for preparation of the graphics.

 
   References Top

1.Koplan BA, Stevenson WG, Epstein LM, Aranki SF, Maisel WH. Development and validation of a simple risk score to predict the need for permanent pacing after cardiac valve surgery. J Am Coll Cardiol 2003;41:795-801.  Back to cited text no. 1
[PUBMED]    
2.Nardi P, Pellegrino A, Scafuri A, Bellos K, De Propris S, Polisca P, et al. Permanent pacemaker implantation after isolated aortic valve replacement: Incidence, risk factors and surgical technical aspects. J Cardiovasc Med 2010;11:14-9.  Back to cited text no. 2
    
3.Limongelli G, Ducceschi V, D'Andrea A, Renzulli A, Sarubbi B, De Feo M, et al. Risk factors for pacemaker implantation following aortic valve replacement: A single centre experience. Heart 2003;89:901-4.  Back to cited text no. 3
[PUBMED]  [FULLTEXT]  
4.Lehr EJ, Wang W, Mullen JC, MacArthur R, Ross D, Koshal A, et al. Abstract 2467: Stentless aortic valve replacement increases the requirement for permanent pacemaker. American Heart Association Scientific Sessions 2007. Circulation 2007;116:II_543-a.  Back to cited text no. 4
    
5.Deeb DM. Aortic Valve Replacement with the medtronic freestyle xenograft using the subcoronary implantation technique. Op Tech Thor Cardio Surg 2006;11:173-84.  Back to cited text no. 5
    
6.Gregoratos G, Abrams J, Epstein AE, Freedman RA, Hayes DL, Hlatky MA, et al. ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices-summary article: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to Update the 1998 Pacemaker Guidelines). J Am Coll Cardiol. 2002;40:1703-19.  Back to cited text no. 6
[PUBMED]  [FULLTEXT]  
7.Kunadian B, Vijayalakshmi K, Thornley AR, de Belder MA, Hunter S, Kendall S, et al. Meta-analysis of valve hemodynamics and left ventricular mass regression for stentless versus stented aortic valves. Ann Thorac Surg 2007;84:73-8.  Back to cited text no. 7
[PUBMED]  [FULLTEXT]  
8.Borger MA, Carson SM, Ivanov J, Rao V, Scully HE, Feindel CM, et al. Stentless aortic valves are hemodynamically superior to stented valves during mid-term follow-up: A large retrospective study. Ann Thorac Surg 2005;80:2180-5.  Back to cited text no. 8
[PUBMED]  [FULLTEXT]  
9.ElKhally Z, Thibault B, Staniloae C, Theroux P, Dubuc M, Roy D, et al. Prognostic significance of newly acquired bundle branch block after aortic valve replacement. Am J Cardiol 2004;94:1008-11.  Back to cited text no. 9
    
10.Fukuda T, Hawley RL, Edwards JE. Lesions of conduction tissue complicating aortic valvular replacement. Chest 1976;69:605-14.  Back to cited text no. 10
[PUBMED]  [FULLTEXT]  
11.Erdogan HB, Kayalar N, Ardal H, Omeroglu SN, Kirali K, Guler M, et al. Risk factors for requirement of permanent pacemaker implantation after aortic valve replacement. J Card Surg 2006;21:211-5.  Back to cited text no. 11
[PUBMED]  [FULLTEXT]  
12.Dawkins S, Hobson AR, Kalra PR, Tang AT, Monro JL, Dawkins KD. Permanent pacemaker implantation after isolated aortic valve replacement: Incidence, indications, and predictors. Ann Thorac Surg 2008;85:108-12.  Back to cited text no. 12
[PUBMED]  [FULLTEXT]  
13.Totaro P, Calamai G, Montesi G, Barzaghi C, Vaccari M. Continuous suture technique and impairment of the atrioventricular conduction after aortic valve replacement. J Card Surg 2000;15:418-22.  Back to cited text no. 13
[PUBMED]    
14.El-Hamamsy I, Clark L, Stevens LM, Sarang Z, Melina G, Takkenberg JJ, et al. Late outcomes following freestyle versus homograft aortic root replacement - results from a randomized trial. J Am Coll Cardiol 2010;55:368-76.  Back to cited text no. 14
[PUBMED]  [FULLTEXT]  
15.Elahi M, Usmaan K. The bioprosthesis type and size influence the postoperative incidence of permanent pacemaker implantation in patients undergoing aortic valve surgery. J Interv Card Electrophysiol 2006;15:113-8.  Back to cited text no. 15
[PUBMED]  [FULLTEXT]  
16.Elahi MM, Osman KA, Bhandari M, Dhannapuneni RR. Does the type of prosthesis influence the incidence of permanent pacemaker implantation following isolated aortic valve replacement. Heart Surg Forum 2005;8:E396-400.  Back to cited text no. 16
[PUBMED]  [FULLTEXT]  
17.Huynh H, Dalloul G, Ghanbari H, Burke P, David M, Daccarett M, et al. Permanent pacemaker implantation following aortic valve replacement: Current prevalence and clinical predictors. Pacing Clin Electrophysio 2009;32:1520-5.  Back to cited text no. 17
    
18.Dagenais F, Cartier P, Dumesnil JG, Pibarot P, Lemieux M, Raymond G, et al. A single centre experience with the Freestyle bioprosthesis: Midterm results at the Quebec Heart Institute. Semin Thorac Cardiovasc Surg 2001;13:156-62.  Back to cited text no. 18
[PUBMED]    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2], [Table 3]



 

Top
  
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
  • Song Z
  • Lehr EJ
  • Wang S
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Acknowledgment
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed839    
    Printed62    
    Emailed0    
    PDF Downloaded95    
    Comments [Add]    

Recommend this journal